Note: In case of supply voltage below 100V, the lock

of the breaker may not latch and the pump might re-

start on its own after sufficient cooling. Take appropri-

ate precautions, if an automatic restart of the pump

may lead to a dangerous situation.

A warm up period (approximately 15 min.) is

required to ensure that the rated ultimate vac-

uum and pumping speed are attained.

O

➨ I

m

3

/h

Torr/mbar

~15min

Shutdown & storage

The pump can be switched off under vacuum.

Condensate?

O

➨ I

(5-10min)

atm

Short-term:

- Has the pump been exposed to condensate?

Allow the pump to continue to run at atmo-

spheric pressure for a few minutes.

- Has the pump been exposed to media which

may damage the pump materials or form

deposits?

Check and clean pump heads if necessary.

Long-term:

- Take measures as described above regard-

ing short-term shutdown.

- Separate the pump from the application.

- Close inlet and outlet ports (e.g., with trans-

port caps).

- Store the pump under dry conditions.

NOTICE

NOTICE

air

Make sure the fan’s air supply is adequate.

Check fan regularly for dust/dirt. Clean fan

guard grill if necessary to avoid a reduction of

ventilation.

NOTICE