Gas inlet temperatures – VACUUBRAND MV 10 User Manual

Page 24

Advertising
background image

page 24 of 51

Gas inlet temperatures

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

Type

ME 16

MD 8

Maximum pumping speed

50/60 Hz (ISO 21360)

cfm

(m

3

/h)

7.1 / 7.6

(12.0 / 12.9)

3.8 / 4.4

(6.5/ 7.5)

Ultimate vacuum (absolute)

Torr

(mbar)

< 60

(< 80)

1.5

(2)

Maximum rated current at:

100 V~ 50/60 Hz***

120 V~ 60 Hz***

230 V~ 50/60 Hz***

400 V~ 50 Hz

A

A

A

A

6.4 / 6.2

5.3

2.8 / 2.7

1.2

Dimensions L x W x H

approx.

in

(mm)

18.5 x 8.7 x 11.6

(470 x 222 x 294)

*** During the start-up period of the pump (first 6 minutes after switching-on) the current draw

might be elevated (up to twice the given nominal current draw).

Type

MD 12

MV 10

Maximum pumping speed

50/60 Hz (ISO 21360)

cfm

(m

3

/h)

5.7 / 6.1

(9.6 / 10.4)

4.8/ 5.2

(8.1/ 8.8)

Ultimate vacuum (absolute)

Torr

(mbar)

1.5

(2)

0.45

(0.6)

Maximum rated current at:

100 V~ 50/60 Hz***

120 V~ 60 Hz***

230 V~ 50/60 Hz***

A

A

A

-

5.3

2.8 / 2.7

6.4 / 6.2

5.3

2.8 / 2.7

Dimensions L x W x H

approx.

in

(mm)

19.1 x 8.7 x 11.6

(486 x 222 x 294)

We reserve the right for technical modification without prior notice!