Cleaning and inspecting the pump heads

Loosen the union nuts (N) of the hose con-

nections at the pump heads with an open-

ended wrench (w/f 15).

Loosen the fittings (Q) at the pump heads

with an open-ended wrench (w/f 17).

Turn the fittings (Q) with an open-ended

wrench (w/f 20) to detach the hoses from

the pump heads.

+

Pull off the one curved connecting hose (R)

at the outlet side of the pump (only MD 12 /

MV 10) directly at the pump head.

Disassemble the housing cover (A) to check

the valves (C).

Unscrew four Allen screws with a 5mm wide

Allen key. Remove the housing cover (A) to-

gether with head cover (D), valves (C) and

O-rings (B, only MD 8 / MD 12 / MV 10).

+

Never use a pointed or sharp-edged tool to remove parts (e.g., screw-

driver). We recommend to use a rubber mallet or compressed air (to

be blown carefully into port).

Remove the head cover (D) carefully from the housing cover (A). Note

position and alignment of valves (C). Remove the valves.

+

Replace valves or O-rings if necessary.

+

Use petroleum ether or other industrial solvent to remove deposits. Do

not inhale vapors.

+

Service only one side of the pump at a time to avoid the mixing of

parts.

In case, position the pump on edge (place pump onto inlet distributor

or outlet distributor respectively). Support the pump appropriately.