VACUUBRAND MV 10 User Manual

Page 42

Advertising
background image

page 42 of 51

(1)

(2)

Put on head cover (D).

+

Pay attention to the correct orientation of the head covers: Align the

nib at the head cover (D) with the notch of the housing cover (A).

+

Make sure that the diaphragm stays centered with respect to the bore

so that it will become clamped uniformly between housing (J) and head

cover (D).

Put the valves (C) an O-rings (B, only MD 8

/ MD 12 / MV 10) in place. See figure for the

correct position of the valves:

Inlet side of pump head: The valve tongue

points at the kidney-shaped orifice in the

valve seat (1).

Outlet side of pump head (2): The valve is

oriented the same direction as the valve at

the inlet side.

Screw in the Allen head screws at the head

covers diagonally at first slightly with a 5

mm wide Allen key, then tighten.

+

Recommended torque: 8.9 ft

.

lb

f

(12 Nm).

Individual performance check of a pump head:

Check the performance of an individual pump head by measuring the

vacuum at its inlet port.

Use a suitable vacuum gauge (e.g., DVR 2) and make sure that it is cor-

rectly calibrated. A vacuum of less than 82 Torr (110 mbar) should be

indicated.

+

If the reading is higher than 82 Torr (110 mbar), recheck the pump

chamber. Make sure that the valves and the diaphragms are correctly

seated (diaphragms concentric with bore).

Check all eight pump heads in the same way, and replace

diaphragms and valves as necessary!

Advertising