The pumps are

not approved for operation below

ground.

Do not operate the pump below ground.

Do not pump substances which may form depos-

its inside the pump. The pumps are not suitable for

pumping substances which may form deposits inside

the pump. Deposits and condensate in the pump may

lead to increased temperatures even to the point of

exceeding the maximum permitted temperatures.

• Check the inlet and outlet of the pump, if there is a

danger of forming

deposits inside the pump, e.g., in

the pump chambers (the pump chamber is the part

between diaphragm and head cover. See section “Re-

placing diaphragms and valves”, pg. 46). Inspect the

pump chambers regularly and clean if necessary.

Consider interactions and chemical reactions of

the pumped media. Ensure that the materials of the

pump’s wetted parts are compatible with the pumped

substances, see section “Technical data”, pg. 24.

When changing the substances pumped, we recom-

mend purging the pump with air or inert gas prior to

changing the pumped media. Purging the pump will

pump out residues and it will reduce the possibility of

reactions of the pumped substances with each other

and with the pump’s materials.

Safety during operation

Adopt suitable measures to prevent the release of dan-

gerous, toxic, explosive, corrosive, noxious or pollut-

ing fluids, vapors and gases. To prevent any emission

of such substances from the pump outlet, install an

appropriate collecting and disposal system and take

protective action for pump and environment.

You must take suitable precautions to prevent any for-

mation of explosive mixtures in the pump chamber or

Advertising