Notice – VACUUBRAND RC 6 User Manual

Page 10

Advertising
background image

page 10 of 59

nected to the pump, to the gas ballast or to a venting

valve.

Note: Flexible elements will shrink when evacuated.

• Connect hoses gas tight at inlet and outlet of the pump.

• Ensure that no foreign objects can be drawn into the

pump.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

• Use only oil of the recommended type. Other oils or

operating fluids may cause damage of the pump or

danger.

Use

special oils (see “Oil change”, pg. 35) for the

rotary vane pump, if, e.g.,

- acid vapors are pumped.

- extremely high aging stability is required.

- oxygen or other strong oxidizing agents are pumped.

• Take adequate precautions when handling pump flu-

ids, lubricants, and solvents. Use appropriate protec-

tive clothing, safety goggles and protective gloves to

avoid excessive contact with the skin and possible

skin irritations (including dermatitis). Do not inhale or

swallow! Ensure that the pump location is well venti-

lated and that possible toxic effects of certain vapors

are avoided. Maintain adequate levels of hygiene and

cleanliness.

• Comply with all relevant statutory requirements and

regulations concerning the handling, storage and dis-

posal of oil.

Note: Do not allow oils to enter the drainage system

or other bodies of water. Spillage can cause accidents

(slip hazard)!

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

NOTICE

Advertising