Notice – VACUUBRAND RC 6 User Manual

Page 26

Advertising
background image

page 26 of 59

• Install a cold trap in front of the pump, if pumping ag-

gressive or corrosive gases or vapors.

• A power failure may cause accidental ventilation of the

pump or the vacuum system, if the manual gas ballast

valve is open. If this constitutes a potential source of

danger, take appropriate safety measures, e.g., install

a solenoid operated gas ballast valve (see “Accesso-

ries”, pg. 39).

• The pump is supplied dry (i.e., without oil filling).

Before starting the pump for the first time, pour in

the oil as described in the above section ”Initial opera-

tion / Pouring in the pump oil”, pg. 24.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Keep a minimum distance of 2 in (5 cm) between the cool-

ing fan and surrounding items (e.g., housing, walls, etc.),

or else install an external automatic ventilation system.

Use connecting hoses with large diameter and keep them

as short as possible to avoid flow losses. Locate the pump

as closely as possible to the application. Use only hoses

at the inlet and outlet of the pump with an inner diameter

at least as large as the diameter of the pump’s connec-

tions.

Always install outlet tubing descending from the pump or

provide other measures to avoid backflow of condensate

towards the pump.

Use a suitable valve (see”Accessories”, pg. 39) to iso-

late the pump from the vacuum application. This is to allow

the pump to warm up before pumping condensable vapors

and to clean the pump after use before it is switched off.

When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

NOTICE

Advertising