Initial operation / Pouring in the pump oil

The pump is supplied dry (i.e., without oil filling).

This is in order to ensure that oil cannot make its way

from the rotary vane pump into the housing of the oil

mist filter during transport. A can containing 0.5 liter

(0.53 quarts) of rotary-pump oil is supplied with the

pump.

Before starting the pump for the first time,

pour in the oil as described below.

Pouring in the oil:

Remove the oil inlet plug (A) (6 mm

wide Allen key).

+

Note: Because new oil contains

gas, a considerable degree of foam-

ing can occur during operation with

new oil, and oil could penetrate the

oil separator.

Pour only approx. 350 ml (0.37 quarts) of oil into the

pump (bringing oil level to ”min.” mark (B), approx.).

Screw in oil inlet plug and with inlet line closed, oper-

ate the pump for approx. 5-10 minutes.

Remove the oil inlet plug again.

Add approx. 100 ml (0.11 quarts) of oil.

+

As a rule, it is not necessary to use all of the oil sup-

plied.

Do not overfill!

Screw in the manometer (C) with its

adapter carefully (open end wrench

width 17 mm).

+

Pay attention to the O-ring!

+

Do not turn the manometer itself as

leaks might be caused!