Write commands compatible to ”cvc 2000 – VACUUBRAND VNC 2 E User Manual

Page 29

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999199 / 28/11/2012

page 29 of 41

Function

operation mode

preset vacuum

preset vacuum

with venting*

preset speed

pressure for restart

delay

pressure for switch off

time for switch off

starting process

control

stopping process

control

remote operation**

driving

venting valve

* Pressure setting with venting is only possible in operation mode ”Vacuum control”, if a venting valve

is connected and configurated and ”Vacuum control” is started. The venting valve opens automati-

cally, if the actual vacuum is 10 mbar below the preset vacuum. Automatic venting becomes inactive,

if ”Vacuum control” is stopped (”Esc”), setting a pressure value using the command OUT_SP_1 or if

the operation mode is changed. Activate the command OUT_SP_V again if necessary.

** If remote operation is selected or deselected, the user has to ensure that no dangerous status of

the system due to the change of the mode of operation can occur and to provide appropriate safety

measures, especially if selecting remote operation interferes with a locally operated active process.

Description

0: VACUULAN

1: continuous pumping

2: vacuum control

4: RC 5/6 management

unit according to preselection (0001 to 1060 mbar

(hPa) or 0001 to 0795 Torr)

unit according to preselection (0001 to 1060 mbar

(hPa) or 0001 to 0795 Torr)

speed in Hz (99.9 = ”HI”)

unit according to preselection (0001 to 1060 mbar

(hPa) or 0001 to 0795 Torr)

hh:mm (hours:minutes)

unit according to preselection (0001 to 1060 mbar

(hPa) or 0001 to 0795 Torr)

hh:mm (hours:minutes) (00:00 = switch off)

1: termination of process control

2: termination of process control and storage of

the actual pressure as new set point

0: remote off

1: remote on

0: close venting valve

1: open venting valve (process control stopped)

(valve does not close again automatically)

store settings permanently

Parameter

X

XXXX

XXXX

XX.X

XXXX

XX:XX

XXXX

XX:XX

X

X

X

Command

OUT_MODE

OUT_SP_1

OUT_SP_V

OUT_SP_2

OUT_SP_3

OUT_SP_4

OUT_SP_5

OUT_SP_6

START

STOP

REMOTE

OUT_VENT

STORE

Write commands compatible to ”CVC 2000”

Attention: If con-

trol is running only

switching from 1

to 2 and 2 to 1 is

possible.