PC6000 Menus

Section 4

YSI Incorporated

PC6000 Software Operations Manual

15

Example:

12/31/92 23:59:59 100
1/15/93 16:02:00 98
2/15/93 18:00:00 94

In this example, an independent DO reading taken at 4:02 PM on 1/15/93 showed a value
of 98 % air-saturation.

3.

After the last line has been typed and entered, hold down Ctrl and press Z, then press
Enter.

4. The message “1 File(s) copied” should appear.

If only one data point is entered into the compensation file, all data after that date and time will be
corrected. If more than one data point is included in the compensation file, data between points will
be corrected linearly.

To enter the compensation filename, press TAB while the highlight cursor is on Compensate. The
highlight cursor will move to the right. Enter the path and file name.

Example: C:\DO.CMP Enter.

When you enter a compensation filename, the program performs a test load of that compensation
file. A summary is displayed so you can verify the file structure. The summary includes how many
lines were accepted, rejected, or blank, the total number of lines, and the earliest and latest date.

If you compensate a constant used in calculating one parameter, you must also compensate the same
constant if you use it to calculate other parameters. A total of 1000 compensation points are
allowed for any single report or plot.

Select Control Info from the Modify menu. The Control Info menu will be displayed.

Enter title line

SAMPLE.DAT

First sample was at:

02/05/92 13:58:00

Enter start time:

begin

Enter stop time:

end

Last sample was at:

02/06/92 11:42:00

Logging interval was:

60 seconds

Enter averaging interval:

0

X-axis time per screen:

1304

Cross-parameter style:

AVG

Cross-parameter bins:

100 (maximum 2350)

Parameters per plot:

2