Cooper-Atkins 93025 User Manual

Page 2

Advertising
background image

7

2

CALIBRATION PROCESS

93025

SysCal

™ Calibration Kit

93025 TROUBLESHOOTING

TROUBLESHOOTING

1) Unit Does Not Turn On:

Batteries are installed

incorrectly or dead. Check batteries are installed

according to picture on rear label. Replace batteries.
2) Display Shows “Prob”:

Probe is broken (Open

thermocouple. See “Test for Broken Probe”
below).

3) Test for Broken Probe:

Pull the probe connector out

partially to expose the blade contacts. Place a metal

paper clip across connector blades to short contacts.

If instrument displays temperature, the probe is

broken. Replace Probe.
4) Battery Icon Appears on Display:

Low battery.

Replace batteries.
5) Display Shows “bAtt” and Instrument Shuts Off:

Replace batteries.
6) Display Shows “LO” or “HI”:

Temperature measured

is outside measurement range of instrument.

TECHNICAL SUPPORT

Cooper-Atkins® Corporation

11353 Reed Hartman Hwy, Suite 110

Cincinnati, OH 45241-2929

888-533-6900

(1) 35025 Thermocouple instrument

(2) 50219-K probes

(1) ISO 17025 certificate of calibration

(1) Calibration log sheet

(1) Hard carrying case

(48) Calibration labels

(48) Blank labels

PROBE CONNECTORS LOCATED INSIDE COLD ENVIRONMENTS

It is recommended that the probe connector is located outside of any cold environments to

avoid any inaccuracies.
If the probe cord is not long enough to fit outside of the cooler or freezer, it can be left in that

environment during the wait time of 3 to 5 minutes. However, if the probe connector is exposed

to these cold environments, and then plugged into the thermocouple instrument, it needs to

acclimate for approximately 4 minutes. This will compensate for the temperature difference

between the probe connector and the instrument to avoid inaccurate readings.

To avoid taking innaccurate readings, do not leave the thermocouple instrument in these

environments longer than the recommended times explained above.

Probe replacement is no longer necessary as long as the calibration process yields an

accuracy within the stated range, and the calibration process is performed within the actual

in-use environment. If any calibration results demonstrate out of tolerance conditions for

a TempTrak

®

transmitter or probe, the equipment will need to be replaced (refer to current

price schedule). For further information contact your local TempTrak sales representative.

SysCal

™ CALIBRATION PROCESS

The Syscal calibration process was designed to monitor the accuracy of a previously

installed TempTrak® transmitter and its modified* Solid Simulator Probe. This process will

be used with the transmitter and solid simulator probe connected, measured as a unit, and

in place within the use environment.
Calibration of the transmitter and solid simulator is achieved by comparing the documented

readings from the transmitter and solid simulator with the reading from the SysCal™

35025 Thermocouple and the 50219-K Probe. This reading is referred to as the reference

standard.
The Total System Accuracy (TSA) of SysCal is ±0.5°F (±0.3°C). A certificate of calibration

from an ISO 17025 accredited laboratory is also supplied. This accuracy will provide a Test

Uncertainty Ratio (TUR) of 4:1 when used as the reference standard for the TempTrak

transmitter and Solid Product Simulator.

*The modified solid simulator incorporates a secondary hole in the opposite end from where the thermistor sensor is inserted
into the solid simulator. This secondary hole is used to place an insertion probe in the vicinity of the simulator’s thermistor
sensor.

Advertising