Cooper-Atkins 93025 User Manual

Page 3

Advertising
background image

CALIBRATION INSTRUCTIONS

1

Insert the SysCal probe into the calibration port

of the solid simulator.

Close the door to the monitoring area to minimize

temperature fluctuations. To ensure an accurate

reading between the probe and the solid simulator,

allow them to stabilize for 3 to 5 minutes before

recording a temperature.

3

A TempTrak reading transmission is initiated

by pressing the transmit button* on the

TempTrak transmitter (900 MHz shown).

*Depending on the transmitter type, this button may be

accessible from the inside (as above) or from the outside.

5

Seal the calibration port on the solid simulator probe

by placing a blank label (supplied) over the hole.

2

Secure probe by looping the elastic band over

the top of the solid simulator.

The reading from the TempTrak

transmitter is compared to the

reference standard recorded by

the SysCal thermocouple. If that

reading is within ±2.0ºF (±1.1ºC), the

transmitter and solid simulator probe

are considered to be calibrated

within the stated accuracy.

4

Immediately after sending the reading from the

transmitter, take note of the temperature displayed

on the thermocouple. Record this reading in your

data log sheet. Use this as your reference standard.

6

A new calibration label (supplied) is dated and signed

by the technician, and placed on the housing

of the TempTrak transmitter (802.11 Wi-Fi shown).

6

3

93025 SYSCAL SYSTEM SPECIFICATIONS

SYSTEM COMPONENTS

:

(1) 35025 thermocouple and (2) 50219-K probes
SYSTEM RANGE:

-25°F to 180°F (-32°C to 82°C).
SYSTEM ACCURACY

:

±0.5°F (±0.3°C) when probe/instrument are calibrated

together and ambient temperature is between 68°F and

86°F (20°C and 30°C).
AMBIENT TEMPERATURE RANGE:

Operating: 32°F to 122°F (0°C to 50°C)
AMBIENT ACCURACY:

At ambient temperatures between 32°F to 68°F (0°C to

20°C) and 86°F to 122°F (30°C to 50°C), add ±0.1° error

per °F or °C deviation from 68°F and 86°F (20°C and

30°C).
RESOLUTION

:

0.1° / 1° user selectable
DISPLAY:

LCD with 0.5” (13 mm) digit height
BATTERY TYPE:

(2) 1.5V (AAA) batteries (ANSI/NEDA-24F, IEC-R03)
BATTERY LIFE:

1800 hours typical
LOW BATTERY:

Indicated by battery icon on display. When battery is fully

depleted “batt” appears on display and unit will shut off

and not operate.
STORAGE

:

-13°F to 167°F (-25°C to 75°C) (battery removed)
COMPLIANCE/REGULATORY

:

Calibrated in an ISO17025 accredited lab, CE certified,

NSF listed, WEEE marked, Rohs compliant.
TRACEABILITY:

We certify that this instrument has been calibrated to

temperature and/or electrical standards traceable to the

U.S. Department of Commerce National Institute of

Standards and Technology (NIST).