Accu-Chek Aviva Expert User Manual

Page 251

Advertising
background image

243

9

Display Shows Possible Cause(s)

Possible Solution(s)

E-53 Bad Test
Error

j

A meter or test strip error has

occurred.

j

Discard the test strip and repeat the test.

j

Your blood glucose may be

extremely high.

j

If this matches how you feel, contact your

healthcare professional immediately. If it does not
match how you feel, repeat the blood glucose test
and see Chapter 2, Testing Your Blood Glucose. If
this still does not match the way you feel, run a
control test with your control solution and a new
test strip. If the control result is within the
acceptable range, review the proper testing
procedure and repeat your blood glucose test with a
new test strip. If the E-53 error still appears for your
blood glucose test, your blood glucose test result
may be extremely high and above the system’s
reading range.

Contact your healthcare

professional immediately. If the control test result
is not within the acceptable range, see Chapter 3,
Control Testing.

52123_07006322001.indb 243

10/18/13 2:32 PM

Advertising