Display Shows Possible Cause(s)

Possible Solution(s)

HI bG Warning

j

Your blood glucose may be

higher than the measuring
range of the system.

j

If you are experiencing any of the common

symptoms of high blood glucose, contact your
healthcare professional immediately.

j

Treat your high blood glucose as recommended by

your healthcare professional.

j

Consider checking bG, ketones, and insulin.

Insulin Sensitivity

j

Insulin sensitivity is outside of

the acceptable meter range.

j

Check your entries and contact your healthcare

professional to determine the appropriate settings.

Invalid Bolus
Advice Times

j

The acting time value is less

than the offset time.

j

The acting time value must be set equal to or

greater than the offset time. Reset the acting time
value or revise the offset time.

Invalid Date

j

The date entered is invalid

(dates for reminders cannot
be set to occur in the past).

j

Re-enter date.

Invalid Hyper
Values

j

The hyper warning limit value

must be greater than all of
your target ranges in the time
block settings.

j

Reset the hyper warning limit or revise the target

ranges in time blocks and re-enter the hyper
warning limit.

j

Enter a hyper warning limit that is above the target

ranges of your time blocks. See Chapter 5,
Changing Meter Settings.