2.3.2 Panel Mounting the Monitor

1. Panel cutout is 356mm H x 440mm W for landscape installation (Figure 6).
2. Install C

Y

RAQ in panel cutout. Ensure all studs clear cutout opening and Poron

gasket on the rear of the unit is flush against panel (critical for NEMA
compliance).

3

N

OTE

: Steps 3 & 4 require a second person

3. Align studs on the C

Y

RAQ with the holes in mounting plate from the rear of

panel (as seen in Figure 6).

Important!

Do not exceed 30 in/lbs when tightening mounting nuts.
4. Also from rear of the panel, second person must install and tighten the fourteen

10-32 mounting nuts.