4.6.1 Media Sensor Calibration Procedure

The ‘Media Sensor Calibration Procedure’ allows the operator to recalibrate the
selected media sensor (reflective or gap) for a selected media. Follow the steps
below to calibrate your sensor.

Note:

Before calibrating, be sure the media sensor is set for the appropriate
media type, see section 4.4.2, also be sure the printhead is latched down
and the cover is closed.

1. Hold the

button while powering up the printer, continue to hold the button

until the PAUSED light turns off then release. Next press and hold the
button, continue to hold the button until the PAUSED light turns on then
release.

2. Place a sample of your media (with

backing) over the media sensor,
close the printhead, and then press
the

button. The printer will flash

the FAULT light until it has finished
analyzing the material.

3. Place a sample of your media

backing or position the reflective
mark of your continuous media over
the media sensor, close the printhead,
and then press the

button. The

printer will flash the FAULT light
until it has finished analyzing the
material.

4. Remove all media from the media sensor, close the printhead, and then press

the

button. The printer will flash the FAULT light until it has finished

analyzing the media sensor for no media.

5. Once the calibration is complete (all three operations performed) press the

+

buttons simultaneously to accept the current calibration for “test” and

exit the

.

6. Use the

button, (feeds blank label) and the

button, (prints test label) to

test the current calibration.

Note:

If you wish to discard your changes and revert back to the previous
calibration simply turn off the printer before Step 7.

7.

Now you can save your changes and resume

by again pressing

the

+

buttons simultaneously.