Iepe input, 2 ai data format, Ai data format – ADLINK USB-2405 User Manual

Page 39: Figure 3-3, Excitation current for iepe sensor measurement

Advertising
background image

Operation

29

USB-2405

the signal is important. When AC coupling is selected, DC offset
present in the input signal is removed. AC coupling configuration
is indicated when the DC content of the input signals is to be
rejected. AC coupling enables a high pass R-C filter through the
input signal path, with corner frequency (-3dB) about 0.4 Hz.

IEPE Input

For applications utilizing sensors such as accelerometers or
microphones, the USB-2405 provides an excitation current
source.

The excitation current is 2 mA for the IEPE sensors, with DC volt-
age offset generated because of the excitation current and sensor
impedance. When enabling IEPE current sources, the USB-2405
automatically sets input configuration to AC coupling.

Figure 3-3: Excitation Current for IEPE Sensor Measurement

3.2.2

AI Data Format

When an A/D converter is used, properties of the signal to be
measured must be ascertained to determine the channel to be

2mA

Constant

Current

VCC

IEPE ON/

OFF

AI Channel

Advertising