2 auto-calibration, 3 saving calibration constants, Auto-calibration – ADLINK PXI-2208 User Manual

Page 98: Saving calibration constants

Advertising
background image

86

Calibration

5.2 Auto-calibration

Through the DAQ-/DAQe-/PXI-2204/2205/2206/2208 card auto-
calibration feature, the calibration software measures and corrects
almost all calibration errors without any external signal connec-
tions, reference voltage, or measurement devices.

The DAQ-/DAQe-/PXI-2204/2205/2206/2208 card comes with an
onboard calibration reference to ensure the accuracy of auto-cali-
bration. The reference voltage is measured in the production line
through a digital potentiometer and compensated in the software.
The calibration constant is memorized after this measurement. We
do not recommended adjustment of the onboard calibration refer-
ence except when an ultra-precision calibrator is available.

NOTES

• Warm the card up for at least 15 minutes before initiating auto-cal-

ibration.

• Remove the cable before auto-calibrating the card since the DA

outputs are changed during the process.

5.3 Saving Calibration Constants

When auto-calibration is completed, you can save the new calibra-
tion constants to the user-configurable banks in the EEPROM.
The date and the temperature when you ran auto-calibration is
saved with the calibration constants. You can store three sets of
calibration constants according to three different environments
and re-load the calibration constants later.

Advertising
This manual is related to the following products: