Remote Control Commands
76
PM-1100
Product-Specific Commands—Quick Reference
*
These commands are not executed if a data acquisition is in progress. The “Program running” message
will be returned.
SENS
CORR COLL
ZERO
*
Perform null
measurement
OFFS
[MAGN]
<99.999> [DB]
Set offset value
FREQ
*
<value> [HZ]
Set sampling rate
FREQ?
(99.9)
Get sampling rate
FREQ CAT?
(99.9;99.9;...)
List sampling rates
POW
REF?
(+99.999 dBm)
Get reference value
REF
DISP
*
—
Set reference
STAT
*
<0|1>
Set absolute or
relative
STAT?
(0|1)
Get absolute or
relative
UNIT
*
<0|1|2|3>
Set power unit
UNIT?
(0|1|2|3)
Get power unit
WAV
*
<9999> [NM]
Set wavelength
WAV?
(9999 nm)
Get wavelength
SYST
ERR?
Error code
Next error from error
queue?
VERS?
Current version
Identification string?
UNIT
POW
*
<WATT|DB|DBM|
WATT>
Set display unit
Command
Parameter/
Response