Remote Control Commands
86
PM-1100
Product-Specific Commands—Description
MMEMory:ACQuisition:SAMPles
Description This command changes the samples parameter for the
acquisition setup.
Syntax
MMEM:ACQ:SAMP<space><samples>
Parameters The <samples> parameter is the number of samples to be set
for the acquisition setup. If an invalid parameter is entered, the
closest valid parameter will be entered instead.
Note
The duration of the acquisition has a direct effect on the values
that can be set for the number of samples.
D x R = S
where
D = duration parameter set with the MMEM:ACQ:DURA
command (in seconds)
R = any of the possible sampling rates in sec
-1
(get the
complete list with the SENS:FREQ:CAT? query)
S = valid number of samples (MUST be a whole number)
Example
MMEM:ACQ:SAMP 200
MMEMory:ACQuisition:SAMPles?
Description This command returns the current number of samples for the
acquisition setup.
Syntax
MMEM:ACQ:SAMP?
Response
Current number of samples in the “1” format
Example
MMEM:ACQ:SAMP?