Remote Control Commands

86

PM-1100

Product-Specific Commands—Description

MMEMory:ACQuisition:SAMPles

Description This command changes the samples parameter for the

acquisition setup.

Syntax

MMEM:ACQ:SAMP<space><samples>

Parameters The <samples> parameter is the number of samples to be set

for the acquisition setup. If an invalid parameter is entered, the

closest valid parameter will be entered instead.

Note

The duration of the acquisition has a direct effect on the values

that can be set for the number of samples.
D x R = S
where
D = duration parameter set with the MMEM:ACQ:DURA

command (in seconds)
R = any of the possible sampling rates in sec

-1

(get the

complete list with the SENS:FREQ:CAT? query)
S = valid number of samples (MUST be a whole number)

Example

MMEM:ACQ:SAMP 200

MMEMory:ACQuisition:SAMPles?

Description This command returns the current number of samples for the

acquisition setup.

Syntax

MMEM:ACQ:SAMP?

Response

Current number of samples in the “1” format

Example

MMEM:ACQ:SAMP?