Dosing pump installation, Dosing pump prime procedure – Hanna Instruments HI 84529-01 User Manual

Page 30

Advertising
background image

www.

hannainst

.com

30

Operational Guide

Dosing Pump

Installation

To install the dosing pump follow the procedure below:

1.

Extend the plunger on the 5 mL syringe to its maximum volume.

2.

Place the syringe in the dedicated spot on the top of the meter (1).

3.

Arrange the bottom of the syringe into the holder on the pump (2). Once
the syringe is in place lower the barrel until it sits fl ush on the holder.

4.

Put the o-ring and syringe-fi xing nut over the syringe (3) and turn
clockwise to secure it in place (4).

5.

Place the valve on the top of the syringe (5). Ensure it fi ts securely.

6.

Insert the aspiration tube into the left side of the valve (6) and replace
the titrant bottle cap with the attached cap (7).

7.

Insert the dispensing tube into the top of the valve (8).

Prime cycle should be performed:

1.

If you notice there is no titrant in the tip

2.

Whenever the dosing system tubes are replaced

3.

Whenever a new bottle of titrant is used

4.

Before starting a pump calibration

5.

Before starting a series of titrations

The prime cycle is used to fi ll the syringe before starting a set of titrations. Two
rinse cycles of the syringe are shown in the fi gure below. The dispensing tube is
connected to the top of the valve and the aspiration tube on the left side.

Dosing

Pump Prime

Procedure