Ph measurement – Hanna Instruments HI 84529-01 User Manual

Page 41

Advertising
background image

www.

hannainst

.com

41

Operational Guide

If CAL DUE is displayed perform an
electrode calibration.
Options:
Press MENU to access the instrument’s
menu.
Press HELP to view the contextual help
whenever time you need additional
information.
Press STIR to start/stop the stirrer.
Press TITRATOR to enter titration mode.
Press CAL to access the calibration menu.
Press LOG to save the current reading. A
message indicating the free log space will
be displayed for a few seconds.
In order to take pH measurements,
follow the next steps: Submerge the pH
electrode 2 cm (0.8’’) and the temperature
probe into the sample to be tested and
stir gently. Allow time for the electrode
to stabilize. When the reading becomes
stable, the

(unstable measurement)

symbol will disappear.
If the pH reading is less than -2.00 pH or
greater than 16.00 pH, the closest full-
scale value will be displayed blinking.
If measurements are taken successively
in diff erent samples, it is recommended
to rinse the electrodes thoroughly with
deionized or distilled water and then
with some of the next sample to prevent
cross-contamination.

The pH measurements are aff ected by temperature. In order to have accurate
pH measurements, the temperature eff ect must be compensated for. To use the
Automatic Temperature Compensation (ATC) feature, connect and submerge
the HI 7662-T temperature probe into the sample as close as possible to the
electrode and wait for a few seconds. The “ATC” message will be shown on the

pH