Operational Guide

Note: The aspiration tube must be inserted in the titrant bottle. The dosing
tip must be placed over a rinse beaker.
Before starting the prime procedure, make sure you are using the appropriate
titrant solution for the selected range.

To prime the burette, select PRIME option
from TITRATION mode.
Adjust the rinses number by pressing the
‚ and  keys and press START.

The number of syringe rinses can be set
between 1 and 5 (at least three rinses
are recommended to ensure that the air
bubbles are completely removed).

To pause the prime process press
the PAUSE key; to continue press the
CONTINUE key. To stop the prime process
press the STOP key.

Dosing
Pump Prime
Procedure