Hanna Instruments HI 98811 User Manual

Page 15

Advertising
background image

15

Two different modes to print / log data are available:
1.

Timed Logging; samples are stored and printed (if print function is
active) at fixed time intervals. Data are stored in the lots 01 to 16.

2.

Log on Demand; samples are stored and printed (if print
function is active) when the

LOG key is pressed. Data are stored

in the lot 00. It's possible to perform the Log on Demand either
in normal mode or in Timed Logging mode.

It is possible to switch from logging without printing to logging with
printing in two ways:
• set the function code 03 to "On" to enable printing, to "Off" to

disable printing - while not in timed logging mode.

• press

ALT and PAPER to toggle between printer enabled and

printer disabled while in Timed logging mode.

TIMED LOGGING MODE

To start Timed logging, press

ALT and LOG.

"tAG" will be displayed on the upper part of
the LCD and "Id" on the lower part. For
Sample ID code reading an i-Button

®

tag

must be touched by the touch probe on the
back of the meter within a 20 second inter-
val. If a tag is recognized the meter will beep
once and if the printing is disabled the last
six digits of the tag's code will be displayed on the LCD for a few seconds.
Tag reading can be skipped by pressing

LOG (a null Sample ID will

be stored).

Note: Tag reading timeout - after 20 seconds left without reading any

tag, the logging operation will be aborted with a longer beep.

The lot number will be displayed for a few seconds, then the "LOG"
"INTV" symbols will appear on LCD and if
printer is enabled a first set of data will be
printed. The "LOG" "INTV" symbols will be
fixed if printer is enabled and will blink if printer is disabled.
The printout provides the following
information:

a - Lot number
b - Logging interval

P R I N T I N G / L O G G I N G F U N C T I O N S

24.9°C

17.10

18/12/2004

#04

1 m

ID 2*3 3 5 4*2

g

S.0001