PASSWORD PROTECTION

Setting the GLP parameters (calibration alarm time-out, instrument
ID code, time and date) can be password protected. If password is set
to a value different from 0000 (factory setting), the user will be asked
to enter the password.
• Select the desired GLP parameter code.
• Enter the password using the arrow keys.

• Press the

ALT and CFM keys

to confirm.

• If password is incorrect, the meter will return to the function

selection mode without any warning message.

• If password is correct, the meter provides access to the GLP

parameters.

PARAMETER SETTING

• Once the parameter code has been entered, the appropriate

message will scroll across the LCD for a few seconds.

• The current value of the selected parameter on the upper LCD

and the parameter code on the lower LCD will be displayed. The
first digit will blink if the parameter can assume continuous
values. All the digits will blink if the parameter can assume only
a fixed set of values.

• Enter the new value using the

Arrow keys.

• Press

ALT and CFM to confirm the value.

The following table lists the setup codes along with the description of
the specific setup items, their valid values and the factory settings
(default):

Code

Valid values

Default

00 Lot data printing/scrolling

00÷16

00

01 Logging interval

1,2,5,10,15,30,60,120,180 min 1

02 Print lots data summary
03 Printer enable

On(enabled); Off(disabled) On

05 Log on demand delete
06 Timed data delete

ALT

CFM

LOT INFO

FNC

CFM