Hanna Instruments HI 98811 User Manual

Page 17

Advertising
background image

17

low battery condition, the message "bAtt" will appear for a few
seconds on the LCD.
Note: When an external adapter or new batteries are connected, the

printing must be manually enabled in order to return to
logging with printing mode.

LOG ON DEMAND

In measuring or timed log mode, press

LOG to store

the current reading. "tAG" will be displayed on the
upper part of the LCD and "Id" on the lower part.
For Sample ID code reading an i-Button

®

tag must

be touched by the touch probe on the back
of the meter within a 20 second interval. If a
tag is recognized the meter will beep once
and if the printing is disabled the last six
digits of the tag's code will be displayed on the LCD for a few seconds.
Tag reading can be skipped by pressing

LOG (a null Sample ID will

be stored).
Note: Tag reading timeout - after 20 seconds left without reading

any tag, the logging operation will be aborted.

The LCD will display "Stor" and the value will be stored in the lot 00
(log on demand data area). If the print function is enabled, a
printout is also produced providing the following information:

a - Sample ID
b - Date
c - Sample number
d - Time
e - Readings ("----" means out

of range or probe not connected)

Note: When the Log on demand data area is full (300 samples

stored) and the

LOG key is pressed, the sample will not be

stored and the LCD will display "FULL". In this case it is
necessary to delete the Log on demand data to free space.

VIEW LOGGING INFORMATION

If

ALT and LOT INFO are pressed during logging, the meter displays

for a few seconds the current lot and the number of logged samples.
Then the meter returns to normal operational mode automatically.
If

ALT and LOT INFO are pressed while the meter is not logging, the

24.9 °C

S.002

17.25

18/12/2004

ID 2*3 9*4 3 6