19

18

SAMPLE NUMBER

During logging it is possible
to know the running sample
number. Press LOG twice
and the display will show the
number of values stored in
the current log.

LOGGING MODE WITH PRINTING

This function is suggested when an immedi-
ate report of measurement is required in addi-
tion to the data recording into memory.

Press the ALT and LOG keys simultaneously
to enter the logging mode. The log and page
numbers will be displayed for a few seconds
to indicate the correct operational mode.

The printer will print a complete set of data
and the "LOG" symbol will appear on the
bottom left corner of the display.

If no key is pressed for about 5 minutes, the
display turns off. During printing, the display
shows time, interval and “LOG” symbol.

To reactivate the display press TEMP key.

The printout provides the following informa-
tion:

a - Date (DD-MM-YY)
b - Current log number
c - Current sample number (in

that particular log)

d - Printing interval indicator in minutes
e - Time (HH-MM)
f - Temperature value/s
g - Channel # (HI 955302 only)

UP

DOWN

TIME

LOG

ALT

TEMP

LOG

LOG

UP

DOWN

TIME

LOG

ALT

TEMP

LOG

PRINT

To continue logging without printing, press
the ALT and PAPER keys at the same time,
and "LOG" will start blinking.

After approximately 5 minutes of inactivity,
the display will switch off but
the logging function remains
active.

To reactivate the display
press TEMP key.

To restart printing, press again the ALT and
PAPER keys simultaneously.

The printout provides the following informa-
tion:
a - Date (DD-MM-YY)
b - Current log number
c - Current sample number

(in that particular lot)

d - Printing interval in minutes
e - Time (HH-MM)
f - Temperature value(s)

Notes:

While in logging mode,
the interval cannot be
changed. First exit the
loggingby pressing the
ALT and LOG keys, then
set the new interval.

If PRINT is pressed while
in logging mode, a print-
out is produced without
affecting the running
sample number.

CH

°C

LOG

UP

PAPER

ALT

PRINT

TEMP

10-01-96

==07== 0005M

0014 *14.25
#1 25.0 °C
#2 27.3 °C

.
.

b
c

d

a

e
f

10-01-96

==07== 0005M

0014 *14.25
#1 25.0 °C
#2 27.3 °C

.
.

b
c

d

a

e
f