Hanna Instruments HI 955201 User Manual

Page 9

Advertising
background image

17

16

PRINTING / LOGGING

WITH HI 955301 AND HI 955302

To print the displayed val-
ues, press PRINT. This
function can be activated
in normal operation mode
as well as during logging and scanning data
on display. When in measurement mode, the
printout provides the following information:

a - Current sample number
b - Date
c - Time
d - Temperature value(s)
e - Channel # (HI955302)

LOGGING MODE

This function is suggested when remote mea-
surements have to be taken automatically
without the necessity of an operator and for a
long period of time. In this mode data will be
stored directly into memory.

Set the appropriate logging interval (see "Op-
erational Guide" section).

Press the ALT and LOG keys simultaneously
to enter the logging mode. The log and page
numbers will be displayed for a few seconds
to indicate the correct operational mode.

The printer will print a complete set of data
and the "LOG" symbol will appear on the
secondary LCD.

PRINT

Data from the second channel can only be
added if the recording mode is exited and a new
log number is entered.

A blinking "CH" will appear
on the display next to the
channel number not utilized
and/or recorded.

When the meter is in record-
ing mode, the "LOG" indica-
tion is displayed on the sec-
ondary display.

To save battery life, if no key is pressed, the
meter goes to stand-by mode.

To reactivate the display
press ON/OFF.

Notes:

It is recommended to power the instru-
ment with an external supply during re-
cording mode, especially when many print-
outs are going to be taken.

Before proceeding with recording, make
sure there is enough paper for your mea-
surements. When the paper is finished
the meter will not advise the operator and
the printouts could be lost.

It is possible to insert a new paper roll
during recording session (see "Printer
Maintenance" section).

If the PRINT key is pressed while still in
recording mode, a printout is produced
without affecting the running number.

Once in recording mode, the printing inter-
val cannot be changed. First exit the re-
cording mode by pressing ON/OFF, then
set the new interval.

TO STOP RECORDING

In order to quit the recording mode, press the
ON/OFF key.

CH

°C

LOG

CH

°C

LOG

ON/OFF

UP

DOWN

TIME

LOG

ALT

TEMP

LOG

Log number

Page number

CH

°C

LOG

==07== 0005M

0001 *14.25
#1 25.0 °C
#2 27.3 °C

.
.

10-01-96

0001 *10.24

#1 20.0
#2 45.4
#3 -----
#4 -----

.
.
.
.

a

b
c
d

e

Advertising
This manual is related to the following products: