26

27

• Reinsert the cuvette into the instrument.

• Press READ to start the reading. The instrument displays the results

in mg/L of free chlorine.

INTERFERENCES

Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approximately 2

minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO

3

or acidity greater than 150 mg/L CaCO

3

,

the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.

Free Chlorine

Free Chlorine

• Press ZERO key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette.

Powder reagents procedure
• Add the content of one packet of HI 93701-0 DPD reagent.

Replace the cap and shake gently for 20 seconds (or 2 minutes for
seawater analysis).

• Wait for a minute to allow the undissolved reagent to precipitate

and reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait

for 1 minute and press READ. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L of free chlorine.

Liquid reagents procedure
• To an empty cuvette add 3 drops of HI 93701A-F DPD1

indicator and 3 drops of HI 93701B-F DPD1 buffer. Swirl
gently to mix, and immediately add 10 mL of unreacted
sample. Replace the cap and shake gently again.

x 3

x 3