Hanna Instruments HI 83205-2008 User Manual

Page 9

Advertising
background image

16

17

Aluminum

ALUMINUM

SPECIFICATIONS

Range

0.00 to 1.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.02 mg/L ±4% of reading

Typical EMC

±0.01 mg/L

Deviation

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the aluminon method. The reaction between aluminum and reagents

causes a reddish tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93712A-0

Ascorbic acid

1 packet

HI 93712B-0

Aluminon reagent

1 packet

HI 93712C-0

Bleaching powder

1 packet

REAGENT SETS

HI 93712-01 Reagents for 100 tests

HI 93712-03 Reagents for 300 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Aluminum method using the procedure described in the

Method Selection section (see page 12).

• Fill a graduated beaker with 50 mL of sample.

• Add the content of one packet of HI 93712A-0 Ascorbic Acid and mix

until completely dissolved.

• Add the content of one packet of HI 93712B-0 Aluminon Reagent

and mix until completely dissolved. This is the sample.

• Fill two cuvettes with 10 mL of sample each (up to the mark).

10 mL

10 mL

# 1

# 2

HI 83205 offers an interactive contextual help mode that assists the user at any time.

To access help screens press HELP.
The instrument will display additional information related to the
current screen. To read all available data, scroll the text using the

keys.

Press Support functional key to access a screen with Hanna service
centers and their contact details.
Press Accessories functional key to access a page with instru-
ment accessories.
To exit support or accessories screens press ESC and the instrument will
return to the previous help screen.
To exit help mode just press HELP or ESC key again and the
meter will display the last screen the user was in before entering
help mode.

Instrument ID

Option: 0 to 9999.
This option is used to set the instrument’s ID (identification number).
The instrument ID is used while exchanging data with a PC.
Press Modify functional key to access the instrument ID screen. Press the

keys in order to set the desired value.

Press Accept functional key to confirm the value or ESC to return to the
setup menu without saving the new value.

Meter information

Press “Select” functional key to view the Instrument model, firmware
version, language version and instrument serial number.
Press ESC to return to the Setup mode.

HELP MODE

HELP MODE

HELP MODE

HELP MODE

HELP MODE

HELP MODE

Advertising