Hanna Instruments HI 83205-2008 User Manual

Page 24

Advertising
background image

46

47

Iron LR

• Press ZERO key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette.

• Fill another graduated mixing cylinder up to the 25 mL mark

with the sample.

• Add the content of one packet of HI 93746-0 TPTZ reagent,

close the cylinder and shake vigorously for 30 seconds. This is the
reacted sample.

• Fill a cuvette with 10 mL of the reacted sample up to the mark

and replace the cap.

• Insert the sample into the instrument.

• Press TIMER and the display will show the countdown

prior to the measurement or, alternatively, wait for 30
seconds and press READ. When the timer ends the
meter will perform the reading. The instrument displays
concentration in µg/L of iron.

25 ml

10 mL

12

3

6

9

30"

Iron LR

IRON LOW RANGE

SPECIFICATIONS

Range

0 to 400 µg/L

Resolution

1 µg/L

Accuracy

±10 µg/L ±8% of reading

Typical EMC

±1 µg/L

Deviation

Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a

violet tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93746-0

TPTZ Reagent

2 packets

REAGENT SETS

HI 93746-01 Reagents for 50 tests
HI 93746-03 Reagents for 150 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Iron LR method using the procedure described in

the

Method Selection section (see page 12).

• Fill one graduated mixing cylinder up to the 25 mL mark

with deionized water.

• Add the content of one packet of HI 93746-0 TPTZ

reagent, close the cylinder and shake vigorously for 30
seconds. This is the blank.

• Fill a cuvette with 10 mL of the blank up to the mark and

replace the cap.

• Place the cuvette into the holder and close the lid.

12

3

6

9

30"

10 mL

25 ml