Hanna Instruments HI 83099 User Manual

Page 32

Advertising
background image

63

62

• Add 12 drops of the HI 93704-0 reagent to

each cuvette. Replace the caps and shake
gently to mix (about 30 seconds).

• Place the blank (#1) into the holder and close the lid.

• Press

Timer and the display will show the countdown prior to zeroing the blank. The display will show

-0.0-” when the meter is zeroed and ready for measurement.

• Remove the blank.

• Insert the cuvette with the reacted sample (# 2) into the

instrument and close the lid.

• Press

Read to start the reading. The instrument displays concentration in

µµµµµ

g/L of hydrazine.

Hydrazine

# 1

# 2

# 1

# 2

INTERFERENCES

Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines

Hydrazine

HYDRAZINE

SPECIFICATIONS

Range

0 to 400

µ

g/L

Resolution

1

µ

g/L

Accuracy

±4% of full scale reading at 25 °C

Typical EMC

±2

µ

g/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, method

D1385-88, p-Dimethylaminobenzaldehyde method. The reaction between hydrazine
and the liquid reagent causes a yellow tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93704-0

Liquid Reagent

24 drops

REAGENT SETS

HI 93704-01 Reagents for 100 tests
HI 93704-03 Reagents for 300 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

• Select the

Hydrazine method using the procedure described

in the

Method Selection section (see page 12).

• Fill one cuvette up to the mark with 10 mL of distilled water.

• Place the cap, insert the cuvette # 1 into the holder and

close the lid.

• Press the

Blank key to start adjusting the light level. The display will show “Blank done” when the

meter is ready to take a zero measurement.

• Fill a second cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

10 mL

# 1

10 mL

# 2

# 1