Hanna Instruments HI 83099 User Manual

Page 43

Advertising
background image

85

84

INTERFERENCES

Interference may be caused by:
Co

2+

must not be present

Fe

2+

must not be present

Al

3+

above 32 mg/L

Ca

2+

above 1000 mg/L (as CaCO

3

)

Cd

2+

above 20 mg/L

Cl¯ above 8000 mg/L
Cr

3+

above 20 mg/L

Cr

6+

above 40 mg/L

Cu

2+

above 15 mg/L

F¯ above 20 mg/L
Fe

3+

above 10 mg/L

K

+

above 500 mg/L

Mg

2+

above 400 mg/L

Mn

2+

above 25 mg/L

Mo

6+

above 60 mg/L

Na

+

above 5000 mg/L

Pb

2+

above 20 mg/L

Zn

2+

above 30 mg/L

Nickel LR

Nickel LR

• Press

Timer and the display will show a countdown prior to adding reagent C or, alternatively, wait

for 15 minutes. Add one packet of HI 93740C-0 EDTA reagent to each beaker, cap and swirl to mix
until completely dissolved.

• Fill one cuvette up to the mark with 10 mL of the blank.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Fill a second cuvette up to the mark with 10 mL of the reacted sample.

• Insert the second cuvette into the instrument.

• Press

Read to start the reading. The instrument displays the results in mg/L of nickel.

Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings,
add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvette and swirl until turbidity is removed.

10 mL

10 mL