Hanna Instruments HI 83099 User Manual

Page 60

Advertising
background image

119

118

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet of

HI 93750B-0 reagent. Replace the cap and gently mix for
one minute by slowly turning the cuvette upside down.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the

measurement or, alternatively, wait for 2 minutes and press

Read.

When the timer ends the meter will perform the reading. The
instrument displays the results in

mg/L (ppm) of potassium (K).

• Press ▲ or ▼ to access the second level functions.
• Press the

Chem Frm key to convert the result in mg/L of potassium oxide (K

2

O).

• Press ▲ or ▼ to return to the measurement screen.

INTERFERENCES

Interferences may be caused by:
Ammonium above 10 ppm
Calcium above 10000 ppm as CaCO

3

Chloride above 12000 ppm
Magnesium above 8000 ppm as CaCO

3

Sodium above 8000 ppm

Potassium LR

Potassium LR

POTASSIUM LOW RANGE

SPECIFICATIONS

Range

0.0 to 20.0 mg/L

Resolution

0.5 mg/L

Accuracy

±3.0 mg/L ±7% of reading at 25 °C

Typical EMC

±0.5 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93750

A-0

Potassium Reagent

6 drops

HI 93750

B-0

Powder Reagent

1 packet

REAGENT SETS

HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 132.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the COLORED OR TURBID SAMPLES procedure at page 17.

• Select the

Potassium LR method using the procedure

described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of sample, up to the mark.

• Add 6 drops of HI 93750A-0 Potassium Reagent, replace

the cap and swirl the solution.

• Place the cuvette into the holder and close the lid.

10 mL