Hanna Instruments HI 9829 User Manual

Page 40

Advertising
background image

40

7.2.3 pH Calibration Error Messages
The HI 9829 displays a series of messages if an error has occurred during
calibration.
If the meter does not accept a pH calibration point, a short message is displayed
to indicate the possible error source. The following screens are examples:

These are the available messages:
• “Input out of scale”: the pH value is out of range. The pH sensor may

require replacement.

• “Check sensor”: the electrode may be broken, very dirty or the user has

attempted to calibrate the same buffer value twice. This may also occur if a
new pH sensor has been installed and”restore factory calib” has not been
exercised.

• “Wrong buffer”: the displayed pH reading is too far from the selected buffer

value. This is often seen immediately after a buffer calibration has been
completed but before the pH sensor has been moved to the next buffer. Check
if the correct calibration buffer has been selected.

• “Invalid temperature”: the buffer temperature is outside the acceptable range.
• “Wrong buffer” / “Contaminated buffer” / “Check electrode”: the buffer is

contaminated or the sensor is broken or very dirty.

• “Check sensor” / “Clean sensor”: the electrode is broken or very dirty.
• “Wrong” / “Clear old calibration”: erroneous slope condition. These messages

appear if the slope difference between the current and previous calibration
exceeds the slope window (80% to 110%). Press <Restore Factory calib.> to
clear old data and continue the calibration procedure, or press ESC to quit
the pH calibration mode.

Advertising