APPENDIX A - PROBE MAINTENANCE

The HI 7698292 probe maintenance kit includes HI 7042S (electrolyte solu-
tion for D.O. sensor), spare membranes with o-rings for D.O. sensor, a small
brush for cleaning EC , o-rings for sensor connectors and a syringe with grease
to lubricate these o-rings.

General Maintenance
• Inspect all sensor connectors for corrosion and replace sensors if necessary.
• Inspect sensor o-rings for nicks or other damage and replace sensor if neces-
sary. Lubricate only with grease from kit.

Use only the supplied grease as some lubricants can cause the
o-rings to expand or affect the turbidity calibration standards.

• After prolonged storage or cleaning, calibration of the sensors is required.
• After use rinse the probe with tap water and dry it. The pH electrode bulb must

be kept moist. Dry the D.O., EC and EC/Turbidity sensors.

Dry ISE sensors and

return to their storage caps if they will not be used for a period of time.

• Check GLP data under “Status” to ensure the sensor is still functioning

properly.

pH and pH/ORP Sensor Maintenance
• Remove the sensor protective cap. Do not be alarmed if any salt deposits are

present. This is normal with pH/ORP electrodes and they will disappear when
rinsed with water.

••

•• Shake down the sensor as you would do with a clinical thermometer to eliminate

any air bubbles inside the glass bulb.

••

•• If the bulb and/or junction are dry, soak the electrode in HI 70300 storage

solution for at least one hour.

••

•• To ensure a quick response time, the glass bulb and the junction should be kept

moist and not allowed to dry. Store the sensor with a few drops of HI 70300
storage solution or pH 4.01 buffer in the protective cap. Tap water may also be used
for a very short period (few days).

Never use distilled or deionized water to store pH sensors

••

•• Inspect the sensor for scratches or cracks. If any are present, replace the sensor.
• Cleaning procedure: clean the sensor frequently by soaking it for 1 minute in

HI 70670

or HI 70671 cleaning solution. After cleaning soak the sensor in

HI 70300

storage solution before taking measurements.