102105-0 DP-IR MANUAL REV M

Page 24 of 35

FLOW ERROR icon is on

Very slow to purge out

• Filter needs to be changed

• Plugged probe

• Internal leak

• Check filters and replace if
necessary

• Clean probe filter and clear or
replace the tubing
• Check tightness of internal

Hydrophobic filter

Maintenance

In order to maintain the DP-IR in good working condition, the following maintenance should be
performed as indicated:

Maintenance

Item Frequency

Self-Test and Calibration

Daily to ensure the instrument is

functioning

properly

Battery recharge

Recharge to full capacity after each use

Replace dust filter

Daily or more often as needed depending on

conditions

**Note**

There is a hydrophobic filter inside the front cap behind the dust filter. Replace this filter should

it clog due to water or dust. There is a back up hydrophobic filter inside the instrument housing

to prevent water damage to the instrument. The exterior filters should be changed as needed by

the user.

**Note**

Do not operate the instrument without the dust filter or without the hydrophobic filter.

**Warning**

To reduce the risk of ignition of a flammable or explosive atmosphere, batteries

must be changed or recharged only in a location known to be non-hazardous.

**Caution**

Only use

HEATH battery pack P/N 102961-0.