Chapter i, Dp-ir system specifications – Heath Consultants Detecto Pak-Infrared (DP-I) User Manual

Page 7

Advertising
background image

102105-0 DP-IR MANUAL REV M

Page 7 of 35

Detection Method:

Infrared Controlled Interface Polarization Spectrometer

Measurement Range:

0-100,000 PPM

1-100%

Gas

Display

Range: Auto

Ranging:

Manual

Ranging:

PPM:

0-10,000 PPM

scale:

0-10,000

%

Gas:

1-100% %

Gas:

0-100%

Sensitivity:

0-1000 PPM: 1 PPM

1000-10,000

PPM:

5

PPM

1-100%

Gas:

0.5%

Accuracy:

Greater of +/- .5% or +/- 10% of reading (typical, std conditions)

(% Gas on manual mode)

Detection Alarms Modes:

Digital Methane Detection (DMD):

Audible tone when detection threshold is exceeded

Adjustable Alarm Level from 1 to 9000 PPM

Tick:

Continuous audio tone relative to concentration

System Fault Warning:

Audible tone and indication on the display

Self-Test & Calibration:

Built in Self-Test and Calibration function verifies operation and

adjusts calibration for maximum sensitivity. Test gas cell integrated

within

the

instrument.

Compliance:

EN

61326-1

Conducted Emissions Class B

EN 61326-1

Radiated Emissions Class B

EN 61000-4-2 4kV/8kV

ANSI

C63.4

Class

B

EN

61000-4-3

3

V/M

FCC 47 CFR Part 15 Class B

Dust Filter:

Replaceable filter provides dust protection. Easy change out quick

disconnect filter cap.

Display:

Large easy to read backlit LCD (.75” Numeric)

Operating Temperature:

0 to + 122° F (-17 to 50° C) (nominal battery voltage)

Humidity:

5 to 95% RH, non-condensing

Enclosure:

IP54 (Water splash and Dust resistant)

Instrument Weight:

5.6 lbs.

Carrying Case:

13 lbs. empty; 21 lbs. filled

24.5” x 21” x 9”

Power Supply:

Internal rechargeable Li-ion battery or

External 12 VDC car battery with optional power conditioner

Battery Operating Life:

8 hours at 32° F with out backlight on

Battery Charger:

External, in-line, 110 - 240 VAC, 50/60 hertz. 10 hours to 90%

charge
Shoulder Strap:

Single over the shoulder padded strap

Speaker Volume:

108 dBs @ Alarm port (A-fast)

Survey Probe:

Quick connect fitting with locking clasp. Adjustable length from

25”

to

41”

Bar hole Probe:

Std: 20”; Optional 36”

Intrinsic Safety:

Class 1 Division 1 Group D T3

UL

913

MetLab

#E112840

CHAPTER I

DP-IR System Specifications

Advertising