7 gdu net selection, 8 test or calibrate selection – Heath Consultants Gas Delivery Unit (GDUnet) User Manual

Page 43

Advertising
background image

net

3-15

3.3.7 GDUnet Selection

net must have the

net registered on the network, in the ‘Send

To’ section, click the drop down arrow and select the required

net, as illustrated in Fig. 3-17.

Fig. 3-17 GDUnet Selection

3.3.8 Test or Calibrate Selection

‘Send To’ also contains the selection to ‘Test’ or ‘Calibrate’
instruments as the required process.

• The ‘Test’ process is selected to simply apply gas then

record instrument status.
This process does not update instrument calibration
settings in any way.

• The ‘Calibrate’ process is selected to perform a full

calibration of the instrument. Gases are applied and the
instrument settings are automatically adjusted until the
instrument reading is the same as each gas applied.
This process updates stored data including new ‘cal-due’
date.

To select ‘Test’ or ‘Calibrate’ option, select the appropriate ‘radio’
button, highlighted in Fig.3-18.