Configure 1 Screen

* Meter Drive (Input Scale Factor)= (X1, X5, X10, X100, X1,000, X10,000) Should correspond to the
amount of uncorrected gas per revolution of the meter.

* Uncorrected Volume Display Scale Factor= (X1, X5, X10, X100, X1,000, X10,000) Sets the scale
for all Uncorrected volume displays throughout the system (LCD, Display1, Profiler etc.)

* Corrected Volume Display Scale Factor= (X1, X5, X10, X100, X1,000, X10,000) Sets the scale for
all Corrected volume displays throughout the system (LCD, Display1, Profiler etc.)

* Uncorrected Volume= (Manual entry) Sets the reading of the Uncorrected accumulator (Usually this
reading matches the mechanical index reading to ensure proper counting).

* Corrected Volume= (Manual entry) Sets the reading of the Corrected accumulator (Changing this
reading will not affect the hourly recording in the “Hourly Profiler”.

* Uncorrected Volume/ Display 4-7 digits = Sets the amount of digits to be displayed on the LCD.
Otherwise referred to as “Electronic masking” of the LCD.

* Corrected Volume/ Display 4-8 digits = Sets the amount of digits to be displayed on the LCD.
Otherwise referred to as “Electronic masking” of the LCD.