Hvc troubleshooting – Heath Consultants Heath Volume Corrector (HVC) User Manual

Page 26

Advertising
background image

26

HVC Troubleshooting

The following may serve as a quick troubleshooting guide if you encounter problems in operating the
HVC:

Symptom

Diagnosis

Will not log on

Verify that proper com port is selected on computer. (If a Serial to
USB adapter is used, find out the proper com port selection in the
“Device Manager”)

If error “Com not ready to send” check to see that only 1 session of
HVC Host is running. After correcting problem, it may be necessary
to reboot the computer.

Make sure that “Default Sandia=ID” in

“Host Setup” is set to 63.

Check Com cable

Check if pushbutton will scroll unit

Check battery voltage on LCD

Call factory for assistance.

Call factory for assistance

Will not scroll

Check battery voltage

Test scroll switch with ohmmeter

Restart instrument. Remove main power connector and then remove
the lithium jumper on the PCB. Wait 5 minutes, reconnect the main
power first (with good battery), then re-connect the lithium jumper
(No data will be lost after restart).

Call factory for assistance.

Display is blank

Check main & lithium battery voltage

Replace all batteries

Check lithium jumper (Scrape pins with a knife to clean if necessary)

Restart instrument. Remove and replace batteries and reconnect.

Call factory for assistance.

Not responding to
applied pressure

Confirm pressure mode set to AUTOCheck connections of transducer

wires

Check connections of transducer wires

Check “Full Scale” value in poly file. Calibrate column/ “Edit
Calibration Terms”, enter the correct “Full Scale”

Set the pressure calibration to default. -> Calibrate Column -> Edit

Calibration Terms -> click the check box next to Pressure Calibration
“Default”. Note: Recalibrate Pressure after these steps.

Replace transducer and recalibrate

Call factory for assistance