When the Self-Test fails. The following table may help to isolate the problem:

Cause

Corrective Action

Low signal strength or noisy

• Make sure battery is fully charged
• Make sure that the unit is in its proper position, and

all the way down and flat

Initial ppm-m too low or too
high

• Make sure battery is fully charged
• Make sure the unit is in its proper position, all the

way down, and flat

• Verify calibration cell is not broken
• Use the Calibration Override

Laser adjustment failed

• Make sure battery is fully charged
• Make sure that the unit is in its proper position, and

all the way down and flat

• Verify calibration cell is not broken


**Caution**

Should the instrument not pass after several attempts, do not use the instrument for
survey work until the problem is corrected. Contact HEATH for further assistance.

101296-0 RMLD IS MANUAL REV J

Page 24 of 40