Heath Consultants Remote Methane Leak Detector Intrinsically Safe (RMLD-I) User Manual

Page 32

Advertising
background image

Excessive false detections
while scanning at longer
distances

• Scanning too fast

• Alarm detection

threshold set too low


• Scanning at the range

limit of the instrument

• Slow down the

scanning rate. Pause at
the long range and
sweep towards you


• Increase the Alarm

detection threshold


• Move in closer

Excessive false detections
while scanning at closer
distances

• Scanning too fast

• Alarm detection

threshold set too low

• Avoid making abrupt

motions while scanning


• Increase the Alarm

Detection Threshold

Excessive false detections
or loss of sensitivity

• Laser output not

optimized


• Alarm Detection

Threshold set too high
or low for conditions

• Perform Self-Test

procedure


• Check the Alarm

Detection Threshold

Error icon or Warning icon
on continuously

• Low battery

• Moisture condensation

on mirror due to rapid
change in temperature


• Internal component

failure

• Check battery level and

recharge if necessary


• Allow for the

temperature of the
instrument to stabilize


• Note error code and

contact HEATH

Battery indicator does not
show full charge after
charging

• Battery not fully

charged

• Battery level

calibration error


• Charge unit until solid

green light on charger
is on

• Run unit until it shuts

off, then fully recharge
with out interruption

Battery indicator bars are
flashing on and off

• Battery level

calibration error

• Run unit until it shuts

off, then fully recharge
with out interruption



101296-0 RMLD IS MANUAL REV J