9.0 - battery back-Up

Some APC models, including the IQ+ controller, are available with a power safe battery back-up option. When installed, this feature will drive the valve closed
in the event of a power interruption to the APC controller. This is more commonly of interest when throttling gate- or pendulum valves are used, because these
valves typically also serve as the vacuum line isolation valve. The following describes the general functionality and specifications of the battery back-up feature.

Battery Back-up Activation and Operation

The controller software continuously monitors the operating voltage delivered
to it by the internal Power Supply. If that voltage drops below a certain thresh-
old level for more than 50 msec, then the IQ+ disconnects all external power
outputs (such as ±15 VDC), exits its present operational state and drives the
valve closed using battery power. Once the valve is closed, the
IQ+ disconnects itself altogether and shuts down.

Valve Close Speed

The speed at which the valve closes under battery
power is the same as during normal operation.

Resuming Normal Operation

Once normal power is back on, the IQ+ controller will reinitialize the butterfly
valve and be ready for operation after about 30 seconds. However, if the IQ+
controller is configured with the valve initialization safety lock function, then
the

J4

serial command needs to be issued before valve initialization can occur.

Battery Pack Information

Specially packaged 15-cell, 18V 500 mA·H Ni-Cd battery assembly.

Battery Life

Depends on various factors including temperature, number of discharge
cycles, battery age, starting charge and valve type. Generally, however, a
new freshly charged battery should close a typical gate- or pendulum valve
ten times, or more, without being depleted to the point where full
actuation is no longer possible.

Recharge Time

Depends on various factors including battery age, state of discharge and to a
lesser degree temperature. Under normal conditions, a fully depleted battery
should charge completely in approximately 5 hours. Recharging occurs auto-
matically provided that the battery is still capable of being charged.

Battery Status

The battery voltage can be read by issuing the RS232 serial command

Gv

. The

controller will respond with the internal Power Supply voltage and the battery
voltage as measured on the main PCB.

NOTE:

The battery voltage reported may not be a true

indication of the battery’s condition, because the voltage is read

while the battery is in an unloaded state. Nor-Cal recommends

replacing the battery approximately every 24 months, irrespective

of apparent condition or usage history. Please refer to

Appendix I for ordering information.

Always

obtain

replacement battery packs from Nor-Cal Products.

I

ntellIsys

A

dAptIve

p

ressure

C

ontrollers

IQ+ OP-LIT 9/08

NOTE:

If a Slave valve has a problem of any kind, the Master

will indicate this by flashing the fault LED. For all other

conditions, the Master LEDs operate the same as a standard

IQ+ controller.

8.13 - master/Slave Object

(continued)

led

StatuS

Flashing Green

Network OK device online

Green

Network OK connection

established

Flashing Red

Connection timed out

TAbLE

8.13.5 SLAvE NETWORk LED STATUS