6200

Calorimeter Operating Instruction Manual

10. Remove all unburned pieces of fuse

wire from the bomb electrodes;
straighten them and measure their
combined length in centimeters.
Subtract this length from the initial
length of 10 centimeters and multiply
this burned length by 2.3 calories
per cm (for Parr 45C10 Fuse Wire)
to obtain the fuse correction. The
scale on the fuse wire card can be
used to obtain this value directly.

11. Titrate the bomb washings with a

standard sodium carbonate solution
using methyl orange, red or purple
indicator. A 0.0709N sodium
carbonate solution is recommended
for this titration to simplify the
calculation. This is prepared by

dissolving 3.76 grams of Na2CO3 in
the water and diluting to one liter.
NaOH or KOH solutions of the same
normality may be used.

12. Analyze the bomb washings to

determine the sulfur content of the
sample if it exceeds 0.1%.

Methods

for determining sulfur are discussed
in Operating Instructions No. 207M.

13. At the end of the testing period, turn

OFF the calorimeter at the power
switch.


5-4