Fluke 789 User Manual

Page 19

Advertising
background image

ProcessMeter™

Getting Acquainted with the Meter

11

Table 4. Rotary Function Switch Positions for mA Output

No.

Position

Default Function

Pushbutton Actions

OUTPUT
X

Test leads in

SOURCE:

Source 0 % mA

Test leads in SIMULATE:

Sink 0 % mA

% STEP X or W: Adjusts output up or down to the next 25 % step
COARSE X or W: Adjusts output up or down 0.1 mA
FINE X or W: Adjusts output up or down 0.001 mA

 sets output to 0 %
 sets output to 100 %

OUTPUT
Y

monp

Test leads in

SOURCE:

Source repeating

0 % -100 %-0 %

slow ramp (m)
Test leads in SIMULATE:
Sink repeating

0 % -100 %-0 %

slow ramp (m)

J (Blue) cycles through:

Fast repeating 0 % -100 % - 0 % ramp (o on display)

Slow repeating 0 % -100 % - 0 % ramp in 25 % steps (n on

display)

Fast repeating 0 % -100 % - 0 % ramp in 25 % steps ( pon

display)

Slow repeating 0 % -100 % - 0 % ramp (m on display)

Table 5. Rotary Function Switch Position for Loop Supply

No.

Position

Default Function

Pushbutton Actions

mA

LOOP POWER

250
HART

Test leads in SOURCE:

Supply > 24 V loop power, measure mA

J (Blue) cycles through:

250

Ω

series resistor for HART

communication switched in

250

Ω

series resistor switched out