Battery life, Maintenance, General maintenance – Fluke 789 User Manual

Page 39: Calibration

Advertising
background image

ProcessMeter™

Battery Life

31

Battery Life

Warning

To avoid false readings, which could lead to
possible electric shock or personal injury,
replace the battery as soon as the battery
indicator (
b) appears.

Table 12 shows typical alkaline battery life. To preserve
battery life:

Use current simulation instead of sourcing when

possible.

Avoid using the backlight.

Do not disable the automatic power-off feature.

Turn the meter off when not in use.

Table 12. Typical Alkaline Battery Life

Meter Operation

Hours

Measuring any parameter

140

Simulating Current

140

Sourcing 12 mA into 500

Ω

10

Maintenance

This section provides some basic maintenance
procedures. Repair, calibration, and servicing not covered
in this manual must be performed by qualified personnel.
For maintenance procedures not described in this manual,
contact a Fluke Service Center.

General Maintenance

Periodically wipe the case with a damp cloth and

detergent; do not use abrasives or solvents.

Calibration

Calibrate the meter once a year to ensure that it performs

according to its specifications. Contact a Fluke Service

Center for instructions.