Configuration – ABUS Technologies Field Logger Data Acquisition/Recorder User Manual

Page 8

Advertising
background image

ABUS TECHNOLOGIES INC.

8

Field

Logger

7. CONFIGURATION

A diskette with configuration software is included in the Field Logger package or

can be downloaded from our web site

www.abustek.com

. It is also included with the

FieldChart software, also available in the ABUS site.

To install the Configurator, execute the program FL_Setup.EXE distributed in

the diskette. The FieldChart software automatically installs the Configurator. The

Configurator presents 4 pages for parameter configuration: Communication

Channels, Acquisitions, and Diagnosis.


7.1 Communication Page

When running the Configurator, first select the serial port (COM1 or COM2), as referred in the

Communication page (See figure communication page, below). After setting up the communication

port, the user can start the Field Logger configuration.

Communication Page

When an unknown Field Logger is to be configured (no information on baud rate or address),

the Configurator can perform an ‘Automatic Search’ on the Field Logger connected to the PC COM port.

It finds out the baud rate and address actually programmed in the Field Logger. To access this function,

go to the Communication page and then click the button “Search”, in the “Automatic Search” box. The

actual Field Logger communication parameters are shown, enabling the user to change them (for

example, when configuring a network with many Field Loggers). To change the communication

parameters, select new values for Baud Rate and Address and then hit the button “Modify”.

The configuration program requires that only one Field Logger be connected to the PC when

the “automatic search” utility is called.