Caution – BINDER KMF 115 User Manual

Page 67

Advertising
background image

KMF (E5.3) 07/2014

page 67/98

CAUTION

Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H. for

a long period.

Dry the appliance completely before shut-down:

Set the humidity to 0 % r.H. and turn on humidity switch (3).

Set the temperature set point to 60 °C / 140 °F for approx. 2 hours (Manual mode).

Only then, shut down the unit at the main power switch (2) and close the water sup-

ply tap.

Having turned off the unit by the main power switch (2), always close the water supply tap.

If you operate the unit at high humidity and then immediately turn off the unit, the internal wastewater col-

lector may overflow due to the condensate. This may lead to the emergence of water at the unit.

CAUTION

Overflow of the internal wastewater tank due to condensate.
Emergence of water at the unit.

Following high humidity operation, do NOT directly turn off the unit.

Pump off the condensate before shut-down:

Set the humidity to 0 % r.H. and turn on humidity switch (4). Operate the unit for at

least 2 hours.

Only then, shut down the unit at the main power switch (3) and close the water sup-

ply tap.

12.1 Function of the humidifying and dehumidifying system

Humidifying system
The humidifying system is located in the humidity generation module. In a cylindrical container with a vol-

ume of about 2 liters an electrical resistance heating evaporates water. The water content is kept exactly

at the boiling point, and thus steam can be immediately generated in sufficient quantity for rapid humidity

increases or for compensation of humidity losses, e.g. by door openings. Condensation forming on the

outer walls of the useable volume is led through a water drain in the outer chamber into the wastewater

can which is pumped off automatically to the wastewater pipe when required.
Freshwater
You can supply the unit with freshwater via a water pipe or by manually filling a freshwater can (option,

chap. 14.7). You can mount the can on the rear of the unit or place it next to the unit.

In order to ensure accurate humidifying, observe the following points with regard to the

freshwater supply:

Supply pressure 1 to 10 bar when connecting to a water pipe

Water type: deionized (demineralized) water

To ensure humidification during 24 hours even at high humidity set-points with manual wa-

ter supply, we recommend filling the freshwater can (option) at the end of each day.

Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F.