7 external freshwater and wastewater cans (option) – BINDER KMF 115 User Manual

Page 72

Advertising
background image

KMF (E5.3) 07/2014

page 72/98

14.6 Additional flexible Pt 100 temperature sensor (option)

An additional flexible temperature sensor Pt100 allows measuring the temperature of the charging materi-

al by means of an independent measuring system utilizing Pt 100 entry. The Pt 100 sensor’s top protective

tube can be immersed into liquid substances

Pt 100 Sensor
PIN 1: Pt 100
PIN 2: Pt 100
PIN 3: Pt 100





DIN socket (4)

DIN plug


A suitable DIN plug is enclosed.

Figure 21: Pin configuration of the DIN socket (4) in the right lateral control panel

Technical data of the Pt100 sensor:

Three-wire technique

Class B (DIN EN 60751)

Temperature range up to 320 °C / 608 °F

Stainless steel protective tube 45 mm length, material no. 1.4501

14.7 External freshwater and wastewater cans (option)

If no suitable in-house water connection is available, you

can manually supply water by filling the optional external

freshwater can. There is an additional external water can

for the waste water. Volume: 20 liters / 0.71 cu.ft.
The cans are placed in holding devices. You can affix

them directly at the rear of the unit or place them next to

the unit.



Figure 22: Rear view KMF with installed external water

cans (option)

1 2 3 4 5 6

1 2 3 4 5 6

Advertising
This manual is related to the following products: